Product introduction The collection needle rubber sleeve resilience tester evaluates the force meas
E-Mail-Adresse: info@standard-groups.com
Product introduction
The collection needle rubber sleeve resilience tester evaluates the force measured by the
collection tube when it passes vertically through a die hole larger than 0.05 mm outside
diameter of the collection needle tube under test.
Meet the standard
YYT1618-2018 "Disposable Human Venous Blood Sample Collection Needle"
Technical parameter
1. Pressure test range: 0 — 50N, error is not more than ±1%, resolution 0.01N.
2. Test time: 1-999S, any choice, the error is not more than ±0.5S.
3. Test speed: 50mm/min, 80mm/min, 100mm/min error is not more than ±2%.
4. Printing method: Each time can print 8, 10 test data or 13 test data and the average
piercing force and deviation value.
5. Measuring distance: 0 — 50mm, error is not more than ±0.1mm.
Operation method
1. Insert the needle at the butt end into the die hole of the clamping fixture 2mm above
the surface.
2. Start the test device to compress the rubber sleeve.
3. Measure the rubber sleeve resilience at 5mm and 15mm between the needle tube and
the needle base.
4. The test result is 0.5N≤ the rubber sleeve resilience ≤1.5N.
Haftungsausschluss: Die oben genannten Inhalte dienen ausschließlich zur Referenz und Kommunikation zwischen Brancheninsidern und garantieren ihre Genauigkeit oder Vollständigkeit nicht. Gemäß den einschlägigen Gesetzen und Vorschriften und den Vorschriften dieser Website sollten Einheiten oder Einzelpersonen, die verwandte Gegenstände kaufen, gültige Qualifikationen und Qualifikationsbedingungen erhalten.