Tag Closed Cup Auto Flash Point Analyser, ASTM D 56 - Qinsun
Professionelle Prüfgeräte Hersteller
Nationale Beratungshotline:+86 021-64208466
Produkte

Tag Closed Cup Auto Flash Point Analyser, ASTM D 56

Introduction:TESTECH TAG Closed Cup Auto Flash Point Analyzer fit to the requirements of intern

WhatsApp: +86 13816217984

E-Mail :info@qinsun-lab.com

Inhalt

Introduction:

STD Closed Cup Auto Flash Point Analyzer fit to the requirements of international standards ISO2719-88. using micro-computer technology control, automatic electronic ignition without gas supply, POST function, measurements accurate, repeatable, reliable performance, simple operation. Oil, chemicals, power, environmental protection, railways, research and other departments ideal instrument flash point testing.Model:FTech- ASTM D 56Standard:ASTM D 56The regulation of ISO2719-88 requirements

Features:

1.7-inch color touch-screen LCD display, English interactive display interface;

2. PID control algorithms automatically adjust the heating curve according to standards;

3. Equipped with automatic detection chip atmospheric pressure to measure atmospheric pressure.

4. A big size data storage function, can store 300 results for any time historical record;

5. Automatic error correction system, automatic stop when temperature is alarm;

6. Mixing, ignition, detection, alarm, cooling, print is automatically controlled;

7. The testing process automatically lidopenning, ignition, alarm, cooling and printing;

8. Platinum alloy heating wire ignition, stable and reliable, high precision;

9. Can be connected to a laptop via USB interface communication, compatible withWINDOWSXP /WINDOWS7 system, network management

10. Multilingual language operation.

11 ,Warming, openings, fire, alarm, cooling, printing is automatically controlled by system.

12. High-speed signal processors, high-precision analog / digital conversion circuit, stable instrument performance and test accuracy

12. Equipped with high-speed thermal mini-printer, print beautiful, fast, with a print spooling function

Technical Parameters:

1. Display method:Critical digital oxygen 7-inch color touch screen display
2. Measurement:Differential detection type thermocouple
3. measuring range:Room temperature -399 ºC
4. Resolution:0.1ºC
5. Repeatability:National Standards: flash point <104 ºC when tolerance ± 1 ºC, flash point ≥104 ºC when tolerance ± 2 ºC
International standards: flash point ≤110 ºC when tolerance ± 1 ºC, flash point> 110 ºC when tolerance ± 2 ºC
6. Printer:High-speed thermal printer, 56 mm paper width
7. Temperature detection:Import platinum resistance (Pt100Ω)
8. Ignition:Airless electronic ignition source
9. self-test function:Instrument automatic fault diagnosis
10. cooling method:Forced air
11. power consumption:<550W
12. use of the environmentTemperature of 10 ºC -45 ºC, humidity 30-80%
13. power supply:AC220V±10%
14. Dimensions:450*280*290 mm
15. Weight:About 15KG


Haftungsausschluss: Die oben genannten Inhalte dienen ausschließlich zur Referenz und Kommunikation zwischen Brancheninsidern und garantieren ihre Genauigkeit oder Vollständigkeit nicht. Gemäß den einschlägigen Gesetzen und Vorschriften und den Vorschriften dieser Website sollten Einheiten oder Einzelpersonen, die verwandte Gegenstände kaufen, gültige Qualifikationen und Qualifikationsbedingungen erhalten.