This Fabric Induction Electrostatic Tester is suitable for the determination of the electrostatic pr
WhatsApp: +86 13816217984
E-Mail :info@qinsun-lab.com
Application:
This Fabric Induction Electrostatic Tester is suitable for the determination of the electrostatic properties of fibers, yarns, fabrics, carpets, upholstery fabrics and other fabrics or various plate-shaped finished product under laboratory conditions.
The Fabrics Induction Type Electrostatic Tester host consists of a corona discharge device and a probe detector.Using a given high-voltage electric field, the fabric is discharged for a certain time, causing the fabric to induce static electricity, thereby performing electrostatic energy measurement, half-life of static voltage decay, and electrostatic residual amount to detect the electrostatic charging property of the fabric to be tested.
The seriousness of the charging of the fabric depends on the level of the static voltage after high-voltage discharge and on the other hand. The rate of static escaping on the fabric is expressed by the electrostatic half-life, which is an important indicator for testing the electrostatic properties of the fabric.
Features:
1. Large screen color touch screen display operation, Chinese, English interface, menu -style operation;
2. Special designed high -voltage occurrence circuits ensure continuous and linear adjustment within the range of 0 to 10000V. Disted high -voltage voltage values, high -voltage adjustment intuitive and convenient;
3. The high -voltage occurrence circuit adopts a fully enclosed module structure. The electronic circuit achieves high -voltage turnover and turned on, which overcomes the shortcomings of high -voltage occurrence circuits in domestic products that can easily cause contacts. The use of use is safe and reliable;
4. Any attenuation period of static electrode voltage: 1%to 99%;
5. The timing method and fixed pressure method can be adopted separately. The instrument uses the digital table to directly display the instantaneous peak, half -life value (or residual static voltage value) and attenuation time of high voltage discharge. Easy to operate;
Standards:
FZ/T01042 , GB/T 12703.1 , GB/T 33730, ISO18080-1
Technical Parameters:
1. Measurement range of static voltage values: 0 ~ 10kV.
2. The half -life time limit: 0-9999.99 seconds, the error is ± 0.1 seconds.
3. Sample disk speed: 1400 rpm.
4, discharge time: 0-999.9 seconds can be adjusted. (Standard requirements: 30 seconds ± 0.1 seconds).
5. Diring distance between needle electrodes and samples: 20mm.
6. Measurement between test probes and samples: 15mm.
7. Sample and size: 60mm × 80mm.
8. Power: 220V, 50Hz, 100W.
9. Dimensions: 600mm × 600mm × 500mm (L × W × H).
10. Weight: about 40kg.
Haftungsausschluss: Die oben genannten Inhalte dienen ausschließlich zur Referenz und Kommunikation zwischen Brancheninsidern und garantieren ihre Genauigkeit oder Vollständigkeit nicht. Gemäß den einschlägigen Gesetzen und Vorschriften und den Vorschriften dieser Website sollten Einheiten oder Einzelpersonen, die verwandte Gegenstände kaufen, gültige Qualifikationen und Qualifikationsbedingungen erhalten.