BFE tester measures medical mask bacterial filtration efficiency via aerosol challenge method.
E-Mail-Adresse: info@standard-groups.com
Key Specification | Specification Range | Resolution | Max. permissible errors |
A Route Sampling Flow | 28.3L/min | 0.1L/min | Within ± 2.0% |
B Route Sampling Flow | 28.3L/min | 0.1L/min | Within ±2.0% |
Spray Flow | (8~10)L/min | 0.1L/min | Within ±5.0% |
Peristaltic Pump Flow | (0.006~3.0)mL/min | 0.001ml/min | Within ±2.0% |
Front Pressure of A Route Sampling Flowmeter | (-20~0)kPa | 0.01kPa | Within ±2.0% |
Front Pressure of B Route Sampling Flowmeter | (-20~0)kPa | 0.01kPa | Within ±2.0% |
Front Pressure of Spray Flowmeter | (0~300)kPa | 0.1kPa | Within ±2.0% |
Negative Pressure of the Chamber | (-90~-120)Pa | 0.1Pa | Within ±1.0% |
Working Temperature | 0-50℃ | ||
Negative Pressure of the Chamber | >120Pa | ||
Data Save Ability | >500000 sets(Scalable capacity) | ||
High Efficiency Particulate Air Filter Properties | ≥99.995%@0.3μm, ≥99.995%@0.12μm | ||
Median diameter of aerosol generator mass | Average diameter: (3.0±0.3)μm; Geometric Standard Deviation≤1.5 | ||
Double Routes 6-lever Andersen Sampling apparatus | LevelⅠ>7μm; Level Ⅱ: (4.7~7)μm; Level Ⅲ: (3.3~4.7)μm; Level Ⅳ: (2.1~3.3)μm; Level Ⅴ: (1.1~2.1)μm; Level Ⅵ: (0.6~1.1)μm | ||
Size of Aerosol Chamber | 600×80×3mm (Length×Diameter×Thickness) | ||
Total number of positive quality control sampler particles | (2200±500)cfu | ||
Negative Chamber Flow Speed | ≥5m3/min | ||
Size of Main Machine | 1470×790×2100mm (L×W×H) | ||
Power Supply | AC220V±10%, 50Hz | ||
Noise of the Tester | <65dB(A) | ||
Weight | ≈150kg | ||
Power Consumption | <1500W |
Medical mask bacterial filtration efficiency tester, also known as BFE Tester, by using the ratio of the upstream bacterial challenge to the downstream residual concentration of the measured medical mask material, to determine the medical mask material(such as meltblown fabric)bacterial filtration efficiency(BFE).The machine provides a certain flow rate of bacterial aerosols.
Haftungsausschluss: Die oben genannten Inhalte dienen ausschließlich zur Referenz und Kommunikation zwischen Brancheninsidern und garantieren ihre Genauigkeit oder Vollständigkeit nicht. Gemäß den einschlägigen Gesetzen und Vorschriften und den Vorschriften dieser Website sollten Einheiten oder Einzelpersonen, die verwandte Gegenstände kaufen, gültige Qualifikationen und Qualifikationsbedingungen erhalten.