Coatings Shear Fatigue Tester
- Coating material research and development: In the research and development stage of coating materi
WhatsApp: +86 13816217984
Email:info@qinsun-lab.com
Inhalt
Standard
- YY/T 0988.13-2017 (China): Specifies shear fatigue testing protocols for coatings on metallic substrates, including load ratios, frequency, and failure criteria.
- ASTM F1160-17 (International): Defines methodologies for evaluating the mechanical performance of implant-coating interfaces under cyclic loading.
Product Advantages
- Ultra-Precision Loading: Electromagnetic servo motor ensures backlash-free operation with 0.001 N shear force resolution.
- Biorelevant Simulation: Integrated 37°C water bath system replicates physiological conditions for in vitro corrosion testing.
- Universal Fixturing: Self-centering clamps accommodate substrates (10–50 mm diameter) and coatings (1–200 µm thickness).
- High-Cycle Endurance: 10 million cycle capacity with ±0.05% load accuracy.
- Compact Design: Benchtop footprint (500 × 450 × 700 mm) suits laboratory environments.
- Regulatory Compliance: Pre-loaded YY/T 0988.13 protocols with automated data reporting for FDA/CE submissions.
Product Features
- Electromagnetic Servo Actuator: 10 kN shear load capacity with programmable R=0.1–10 ratios.
- Dual Sensor System: 0.01 N resolution shear force transducer + 0.001 mm LVDT displacement sensor.
- Dynamic Load Control: 0.1–30 Hz frequency range with sine/square/random waveforms.
- Thermal-Mechanical Bath: 37±0.5°C recirculating water system with anti-evaporation design.
- High-Resolution Imaging: Built-in 5MP camera system for coating delamination analysis.
- Safety Suite: Emergency stop, overload clutches, and water leak detection system.
Technical Parameters
Parameter | Spezifikation |
---|
Max. Shear Force | ±10,000 N (0.01 N resolution) |
Frequency Range | 0.1–30 Hz (0.01 Hz increments) |
Load Ratio (R) | 0.1–10 (adjustable) |
Displacement Range | ±2 mm (0.001 mm resolution) |
Water Bath Temperature | 37±0.5°C (PT1000 sensor) |
Data Sampling Rate | 10 kHz (synchronized across 8 channels) |
Power Requirements | 220V/110V ±10%, 50/60Hz, 1,500W |
Dimensions (W×D×H) | 500 × 450 × 700 mm |
Weight | 70 kg (net) / 80 kg (with water bath) |
Noise Level | <55 dB (A-weighted) |
Standard Accessories
- Shear Fixture Kit:
- Self-centering clamps (10–50 mm substrate compatibility)
- Coating thickness gauges (1–200 µm measurement range)
- Thermal-Mechanical Bath:
- 15L capacity with recirculating pump
- Corrosion-resistant titanium interior
- CoatingTest Pro Software:
- Automated S-N curve generation and adhesion failure analysis
- 21 CFR Part 11-compliant audit trail with blockchain timestamping
- Remote monitoring via iOS/Android app
- Operator Kit:
- Calibrated torque wrench set (0.1–10 Nm)
- Substrate alignment jigs (ASTM F1160 compliant)
- Calibration Bundle:
- NIST-traceable force calibration weights
- LVDT verification tool
- Safety Kit:
- Water leak detection system
- Acoustic enclosure with 25 dB noise reduction
Haftungsausschluss: Die oben genannten Inhalte dienen ausschließlich zur Referenz und Kommunikation zwischen Brancheninsidern und garantieren ihre Genauigkeit oder Vollständigkeit nicht. Gemäß den einschlägigen Gesetzen und Vorschriften und den Vorschriften dieser Website sollten Einheiten oder Einzelpersonen, die verwandte Gegenstände kaufen, gültige Qualifikationen und Qualifikationsbedingungen erhalten.