- Evaluating the wear - resistance of hip joint prostheses: By simulating human movement states such
WhatsApp: +86 13816217984
Email:info@qinsun-lab.com
Standard
Parameter | Spezifikation |
---|---|
Angular Motion Range | Flexion/Extension: ±100° Abduction/Adduction: ±30° Rotation: ±40° |
Max. Axial Force | ±4,500 N (1,000 N resolution) |
Load Control Accuracy | ±0.5% of reading (0–4,500 N) |
Frequency Range | 0.1–5 Hz (programmable waveforms) |
Temperature Control | 37±2°C (PT1000 sensors) |
Lubrication System | 15L capacity with protein skimmer and bubble trap |
Data Acquisition | 100 kHz sampling rate (16-bit resolution) |
Power Requirements | 220V/110V ±10%, 50/60Hz, 2,500W |
Dimensions (W×D×H) | 1,800 × 1,200 × 2,000 mm |
Weight | 950 kg (including bath and control cabinet) |
Noise Level | <65 dB (A-weighted) |
Haftungsausschluss: Die oben genannten Inhalte dienen ausschließlich zur Referenz und Kommunikation zwischen Brancheninsidern und garantieren ihre Genauigkeit oder Vollständigkeit nicht. Gemäß den einschlägigen Gesetzen und Vorschriften und den Vorschriften dieser Website sollten Einheiten oder Einzelpersonen, die verwandte Gegenstände kaufen, gültige Qualifikationen und Qualifikationsbedingungen erhalten.
Unternehmen Telefon
+86-21-6420 0566
Arbeitszeiten
Montag bis Freitag
Handy:
13816217984
Email:
info@qinsun-lab.com