Detailed Product DescriptionVolumetric Resistivity Measurement Voltage:DC500V±10%Volume Resisti
WhatsApp: +86 13816217984
E-Mail :info@qinsun-lab.com
Volumetric Resistivity Measurement Voltage: | DC500V±10% | Volume Resistivity Range: | 2.5×10 6~2×10 13Ω.m |
---|---|---|---|
Accuracy: | Higher Than ±10% | Resistance Measurement Range:: | 2M~2TΩ |
Temperature-controlled Induction Furnace Maximum Power: | 0.00001~1 | Resolution Of Dielectric Loss Value: | 0.00001 |
Capacitance Measurement Range: | 10.0pF~200.0pF | Resolution Of Capacitance Value: | 0.01pF |
Empty Cup Capacitance: | 60±5pF | Dissipation Value Measurement Accuracy: | ± (1% Reading Value + 0.02%) |
Capacitance Value Measurement Accuracy: | ± (1% Reading Value + 1pF) | Working Power Supply: | AC220V±10%, 50Hz |
Measurement And Control Temperature Range: | Room Temperature ~ 119.9 ℃ | Measurement And Control Temperature Stability: | ± 0.5 |
Relative Humidity: | ≤85% | Dissipation Measurement Voltage: | 1.5kV, 2.0kV, 2.5kV (conventional) |
Ambient Temperature: | -5℃~50℃ | Dimension: | 480mm×400mm×420mm |
Weight: | 25.7kg | ||
Highlight: | oil analysis lab equipment oil analysis equipment |
The automatic oil dielectric loss and volume resistivity tester conforms to GB/T5654 standard and is used to determine the dielectric loss factor and volume resistivity of insulating materials that are liquid at the test temperature.
It is used to determine the dielectric loss factor and volume resistivity of insulating materials in liquid form at test temperature, including insulating liquids such as transformers, cables and other electrical equipment. Can be widely used in
Electric power, petroleum, chemical industry, commodity inspection and scientific research and other departments.
1. using medium frequency induction heating, room temperature heating to control temperature (90 ℃)
And constant temperature automatic measurement only 15 minutes
2. Measure oil dielectric loss and volume resistivity at the same time or choose one
3. Adopt large screen LCD display, just follow the Chinese menu
prompt, enter the command, the instrument can work automatically
4. With communication function, can configure computer for real-time monitoring.
Dynamic observation of oil dielectric loss value with oil temperature changes and depicted as a graph
5. Automatic display of measurement results, and data printing and saving
6. With over-voltage, over-current, short-circuit protection, and high voltage
indicates, also has the alarm prompt function
Volumetric resistivity measurement voltage: DC500V±10%
- Volume resistivity range: 2.5×10 6~2×10 13Ω.m
- Accuracy: higher than ±10%
- Resistance measurement range: 2M~2TΩ
- Dissipation measurement range: 0.00001~1
- Resolution of dielectric loss value: 0.00001
- Capacitance measurement range: 10.0pF~200.0pF
- Resolution of capacitance value: 0.01pF
- Empty cup capacitance: 60±5pF
- Dissipation value measurement accuracy: ± (1% reading value + 0.02%)
- Capacitance value measurement accuracy: ± (1% reading value + 1pF)
- Working power supply: AC220V±10%, 50Hz
- Measurement and control temperature range: room temperature ~ 119.9 ℃
- Measurement and control temperature stability: ± 0.5
- Relative humidity: ≤85%
- Dissipation measurement voltage: 1.5kV, 2.0kV, 2.5kV (conventional)
(2.0kV) (positive connection) - Ambient temperature: -5℃~50℃
- Dimension: 480mm×400mm×420mm
- Weight: 25.7kg
Haftungsausschluss: Die oben genannten Inhalte dienen ausschließlich zur Referenz und Kommunikation zwischen Brancheninsidern und garantieren ihre Genauigkeit oder Vollständigkeit nicht. Gemäß den einschlägigen Gesetzen und Vorschriften und den Vorschriften dieser Website sollten Einheiten oder Einzelpersonen, die verwandte Gegenstände kaufen, gültige Qualifikationen und Qualifikationsbedingungen erhalten.