Detailed Product DescriptionSTD:Closed Cup Flash Point TesterStandard:ASTM D93Oil Cup Inner Dia
WhatsApp: +86 13816217984
E-Mail :info@qinsun-lab.com
STD: | Closed Cup Flash Point Tester | Standard: | ASTM D93 |
---|---|---|---|
Oil Cup Inner Diameter: | 50.7mm~50.8mm. | Oil Cup Depth: | 55.7mm~56.0mm |
The Scribed Line Depth Of Capacity Of Testing Oil: | 33.9mm~34.3mm. | Capacity Of Testing Oil: | About 70ml |
Thermometers: | Scale 20°C~150°C, Division 1°C. | Keywords: | Oil Analysis Equipment |
Name: | Closed Cup Flash Point Tester | Dimension: | 410X2600X300mm |
Highlight: | 70ml Flash Point Tester ASTM D93 Flash Point Tester Closed Cup Flash Point Tester |
Closed Cup Flash Point Automatical Measuring Instrument is mainly used for the determination of closed flash point value of petroleum products. The instrument adopts microcomputer technology, large screen LCD liquid crystal display, Chinese menu, power failure storage, instrument has automatic ignition, display, Locking and printing results, automatic cooling and other functions, with the advantages of accurate measurement, good repeatability, stable and reliable, easy to operate. Widely used in electric power, petroleum, chemical, commodity inspection, scientific research and other departments, in line with ASTM D93. standard method requirements.
1. This instrument is designed as per ASTM D93.
2. The heating power is continuously and sleeplessly adjustable. The power is shown visually by voltmeter. The temperature control mode is advanced and reasonable.
3. The structure is designed small and exquisite. The stainless-steel table board is beautiful and easy to clean.
4. Operation is easy. Test results are precise. It has been chosen as the instrument to test the closed cup flash point by many metering and detecting institutes.
Display | Large screen LCD module display. |
Measuring range | room temperature ~ 300 ° C |
Temperature detection | platinum resistance |
Accuracy | ≥104°C±2°C ≤104°C±1°C |
Ignition mode | electric ignition |
Cooling method | forced air cooling |
Self-test function | test head, igniter, fan, etc |
Standard | in accordance with ASTM D93 GB261-83 ISO-2719 |
Power source | AC 220V ± 11V, 50Hz ± 2.5HZ |
Power | ≤100VA |
Ambient temperature | 10 ° C ~ 35 ° C |
Weight | 15KG |
Use ambient humidity | ≤ 85% |
Haftungsausschluss: Die oben genannten Inhalte dienen ausschließlich zur Referenz und Kommunikation zwischen Brancheninsidern und garantieren ihre Genauigkeit oder Vollständigkeit nicht. Gemäß den einschlägigen Gesetzen und Vorschriften und den Vorschriften dieser Website sollten Einheiten oder Einzelpersonen, die verwandte Gegenstände kaufen, gültige Qualifikationen und Qualifikationsbedingungen erhalten.
Unternehmen Telefon
+86-21-6420 0566
Arbeitszeiten
Montag bis Freitag
Handy:
13816217984
Email:
info@qinsun-lab.com