Detailed Product DescriptionHydrogen Flame Ionization Detector (FID:Cylindrical Collector Struc

WhatsApp: +86 13816217984

E-Mail :info@qinsun-lab.com

Inhalt

Detailed Product Description

Hydrogen Flame Ionization Detector (FID:Cylindrical Collector Structure Design, Metal NozzleDetection limit:≤2×10-12g/s (n-hexadecane/isooctane)
Baseline noise:≤10μVBaseline drift:≤100μV/30min
Linearity:≧105Adjustable fully automatic ignition, stabilization time:30 Minutes
Thermal Conductivity Detector (TCD):Semi-diffused Structure, Constant Current Control Mode For Power SupplySensitivity:≥5000mV·ml/mg
Temperature range:5°C~420°C Above Room Temperature Accuracy ±0.1°CPower supply:220V±22V, 50Hz
Power:≥2kWWeight:55KG
Dimensions:60cm×50cm×50cm
Highlight:

Semi Diffused Structure Gas Chromatograph

Insulating Oil Gas Chromatograph

Product description

It is designed and manufactured according to the methods stipulated in GB/T 17623 and DL/T 703 standards, and is suitable for analyzing oil-filled electrical equipment (including transformers, reactors, current transformers, voltage transformers, charging bushings, etc.) Analysis of hydrogen, carbon monoxide, methane, carbon dioxide, ethylene, ethane, acetylene and other gas content in oil.

Technical features

Realize computer real-time control and data processing:

High-precision, stable and reliable temperature control system:

The heater will stop heating and report the fault location on the display;

Simple and clear man-machine dialogue interface, easy to operate, easy to learn and use

Double stable high-precision air path control system.

The column chamber adopts the method of tracking temperature rise.

The instrument detects low-level hydrocarbons and high-level CO and CO2. which can be detected separately to avoid mutual interference.

Technical parameters

Hydrogen Flame Ionization Detector (FID)Cylindrical collector structure design, metal nozzle
Detection limit≤2×10-12g/s (n-hexadecane/isooctane)
Baseline noise≤2×10-13A
Baseline drift≤2×10-12A/30min
Linearity≥106
Adjustable fully automatic ignition, stabilization time30 minutes
Thermal Conductivity Detector (TCD)Semi-diffused structure, constant current control mode for power supply
Sensitivity≥5000mV·ml/mg
Baseline noise≤10μV
Baseline drift≤100μV/30min
Linearity≧105
Temperature range5°C~420°C above room temperature Accuracy ±0.1°C
Power Supply220V±22V, 50Hz
Power≥2kW
Weight55KG
Dimensions60cm×50cm×50cm

Haftungsausschluss: Die oben genannten Inhalte dienen ausschließlich zur Referenz und Kommunikation zwischen Brancheninsidern und garantieren ihre Genauigkeit oder Vollständigkeit nicht. Gemäß den einschlägigen Gesetzen und Vorschriften und den Vorschriften dieser Website sollten Einheiten oder Einzelpersonen, die verwandte Gegenstände kaufen, gültige Qualifikationen und Qualifikationsbedingungen erhalten.