ASTM D93 Flash Point Oil Analysis Equipment Pensky Martens Closed Cup Flash Point Tester - Qinsun
Professionelle Prüfgeräte Hersteller
Nationale Beratungshotline:+86 021-64208466
Produkte
Aktueller Standort:Home > Prüfgerät > Sonstige Prüfmittel > Sonstige >

ASTM D93 Flash Point Oil Analysis Equipment Pensky Martens Closed Cup Flash Point Tester

Detailed Product DescriptionSTD:Closed Cup Flash Point TesterStandard:ASTM D93.Oil Cup Inner Di

WhatsApp: +86 13816217984

E-Mail :info@qinsun-lab.com

Inhalt

Detailed Product Description

STD:Closed Cup Flash Point TesterStandard:ASTM D93.
Oil Cup Inner Diameter:50.7mm~50.8mm.Oil Cup Depth:55.7mm~56.0mm
The Scribed Line Depth Of Capacity Of Testing Oil:33.9mm~34.3mm.Capacity Of Testing Oil:About 70ml
Thermometers:Scale 20°C~150°C, Division 1°C.Name:Pensky Martens Closed Cup Flash Point Tester
Keywords:ASTM D93 Flash Point Oil Analysis Equipment
Highlight:

oil analysis equipment

oil analysis machine

Description 

This instrument is designed and manufactured as per standards GB/T 261-2008 Determination of Flash Point - Pensky-Martens closed cup method and ASTM D93 Standard Test Methods for Flash Point by Pensky-Martens Closed Cup Tester. It is used to make determination of the closed cup flash point of the petroleum products.

Merkmale

1. This instrument is designed as per ASTM D93.

2. The heating power is continuously and sleeplessly adjustable. The power is shown visually by voltmeter. The temperature control mode is advanced and reasonable.

3. The structure is designed small and exquisite. The stainless-steel table board is beautiful and easy to clean.

4. Operation is easy. Test results are precise. It has been chosen as the instrument to test the closed cup flash point by many metering and detecting institutes.

Specifications

1. Power supply:AC (220±10%)V, 50Hz.

2. Heating device: The heating power is continuously adjustable from 0 W to 600W.

3. Heating rate ::(1~12)°C/incontrollable and adjustable.

Stirring rate

(1) Procedure A: (90~120) RPM

(2) Procedure B: (250±10) RPM

Standard oil cup

(1) Inner diameter: 50.7mm~50.8mm.

(2) Depth: 55.7mm~56.0mm

(3) The scribed line depth of capacity of testing oil: 33.9mm~34.3mm.

(4) Capacity of testing oil: about 70ml

Igniting device

(1) Igniting source: gas(or other civilian fuels, the same below)

(2) Electric ignition. Gas flame diameter: 3.2mm~4.8mm

Thermometers

Mercury-in-glass thermometer. Specifications are as below:

(1)Scale -5°C~110°C, division 0.5°C.

(2)Scale 20°C~150°C, division 1°C.

(3)Scale 90°C~370°C, division 2°C.

8. Ambient temperature:≤ 35°C

9. Relative humidity:≤ 85%

10. Maximum power consumption:650W

Haftungsausschluss: Die oben genannten Inhalte dienen ausschließlich zur Referenz und Kommunikation zwischen Brancheninsidern und garantieren ihre Genauigkeit oder Vollständigkeit nicht. Gemäß den einschlägigen Gesetzen und Vorschriften und den Vorschriften dieser Website sollten Einheiten oder Einzelpersonen, die verwandte Gegenstände kaufen, gültige Qualifikationen und Qualifikationsbedingungen erhalten.